Features

  • Offer variable energy level selection. (2, 3, 5,7, 10, 20, 30, 50, 70, 100, 150, 200 Joules)
  • Battery capacity – 50 shocks on single charge.
  • Adult & paediatric paddles support.
  • Compact & Light weight : Approx.4.5 Kgs.
  • Printer
  • Optional: Non invasive pacing, SPO2 , NIBP, AED.

Technical Specifications

General

Display : 320 x 240 pixels, 5.7” Blue LCD with LED backlit
Display area : 120mm x 90mm
Dimensions : H180 x W230 x D210mm (max.) without Paddle
Weight : 4.5 Kgs.

Defibrillator

Operating Mode : Manual
Waveform : Biphasic
Energy : 2 to 200 joules in 12 steps
Synchronous : Energy delivery begins within 60 ms of the QRS Peak
Charging Time : Less than 8 seconds
Paddle Assembly : External paddle assembly with coil retractable high voltage cable Adults & Pediatric paddle integrated to same handle
Check Mode : Manual test mode
Operation : By front Panel keyboard and paddle buttons

ECG:

Frequency Response : 0.5 Hz to 40 Hz (ECG)
Frequency Response : 0.5 Hz to 30 Hz (Paddle)
Patient isolation : Optical isolation
Lead selection : Lead I,II and III
Gain setting : 0.5, 1.0 and 2.0
CMRR : > 90 db
Sweep Speed : 12.5, 25.0, 50.0 mm/sec
Waveform 1 : ECG
Waveform 2 : 4sec delayed (cascaded)
Data Output : Analog output of ECG
Digital filter : Digital signal processing for good quality of ECG
Freeze mode : To freeze last 4 seconds ECG

Heart Rate

Range : 30-250 BPM
Accuracy : +/- 2 BPM
Resolution : 1 BPM
Tone : Beep on every QRS detected
Low alarm Limit Range : 30-245 (step of 5)
High alarm Limit range : 40-250 (step of 5)

Printer

Paper Type : Thermal paper 50/55 mm width, 20 meters length
Event Printing
ECG Mode : Prints 5 Seconds pre & post ECG
ECG in Defib
Mode : Prints 5 Seconds pre & post ECG
Defib Mode : Stores & prints 10 Seconds of Defib Shock data i.e. 5 seconds pre critical & 5 seconds post critical Maximum events that can be store Memory of 10
Manual Defib
Check Mode : Prints the status of Check Mode
Print Annotates : Time, Date, ECG Lead, Mode, Shock Status & selected Energy
Delay : 10 Seconds

Power Supply (AC Mains Through Adaptor)

Input : 150-276VAC, 50Hz, 60Hz, 12V DC for ambulance
Output : 17V@4A
Battery : 12V/3.8 Ah Rechargeable
Back-up time : 3 hours min on full charge in ECG mode /50 shock of 150 Joules in defib mode
Power consumption : 30 Watts (max) in ECG mode

Optionals

Non Invasive Pacing
SPO2
NIBP
AED

Environmental Conditions

Operating Temp. : (-)5 ⁰C ~ (+)50 ⁰C
Heart Rate Humidity : Up to 95 % non-condensing