Meditec-1700_cat

Meditec 1700 Ventilator

 

Features

  • Invasive as well as Non-Invasive Ventilation
  • Pressure & Flow waveforms with Trends Facility
  • Altitude compensation
  • FiO2 Measurement
  • Suitable with both dual as well as single limb breathing circuits
  • Patient History storage capability
  • Automatic Parameters adjustment of patient according to Height & Weight
  • Battery Backup: 3 Hrs
  • Extended battery backup (optional)
  • Touch screen as well as knob operated
  • Shock Resistant (Drop Test compliant)
  • Optional: EtCO2 & SpO2 Measurement
  • Nebulizer: 10–180 minutes
  • Inspiration hold: Measures patient’s lung compliance and resistance, Elasticity, Time constant
  • Key lock: Key & LCD Touch Screen Lock
  • Waveform freeze: Freezes the waveform
  • Alarm silence: Mutes audible alarms for 2 minutes
  • Alarm reset: Clears visual indicators and messages

The Meditec 1700 is designed to work in a highly versatile environment.Whether you need Invasive or Non-Invasive ventilation the Meditec 1700 delivers.
Power packed with a range of Ventilation modes to suit individual preferences the Meditec 1700 is ideal for use in an ICU, Intra-Hospital transport, Air and Ground transport.
Simplicity, In-built air supply, versatility and affordable cost of ownership make Meditec 1700 the obvious choice..

Technical Specifications

FeaturesSpecifications
Display Screen8 Inch Screen with Touch Control & Rotary Knob
ParametersSet Parameters / Measured Parameters / Alarm Status / I:E Ratio
WaveformPressure-Time, Flow-Time, Volume-Time
LoopPressure-Volume, Flow-Volume, Minute Volume
ModesPACV, PSIMV, VACV, VSIMV, AUTO (Emergency Quick Start), Spont+, PRVC,
Tidal Volume50 - 2500 ml
Inspiratory pressure5 - 80 cm H2O (Peep+Pins < 100cm H2O)
Pressure support0 - 60 cm H2O (PEEP+PSV < 100cm H2O)
PEEP/CPAP0 - 50 cm H2O
Respiratory rate2 - 80 bpm
Inspiratory time0.1 - 24 sec
Rise Time (in PCV)Fast, Medium, Slow (0 - 2 sec)
Apnea Time2 - 60 sec
Apnea BackUp TypePACV , VACV Mode
Trend72 hrs
Enable inhalation trigger(En-sense)10 - 80% of Inhalated volume
Exhalation trigger Sensitivity(Ex-trig)10 - 80% of peak flow in exhlation
Inspiratory Pressure limit5 - 80 cm H2O in PRVC
Trigger sensitivityPressure trigger : OFF, 0.5 to 20 cm H2O
Flow trigger : OFF, 0.5 to 20 lpm
FiO2%21% - 100%
Suitable Body WeightVolume Mode: 10 - 150 kg
Pressure mode: 1 - 150 kg
Nebulizer (time)10 - 180 mins (3 Hours)
Altitude Volume Compensated(-)100 - 5000 [m]
Room Air Temperature0 - 50°C
AlarmsAirway Pressure / Tidal Volume / Minute Volume / Respiratory Rate / Apnea /
O2 % / Airway Leak / Circuit Open / Battery Indicator / O2 Supply Pressure Failure
/ Ventilator In-operative
PowerAC: 100 - 230v 50 - 60 Hz, DC :10.8v - 8.8A
Power Backup180 mins. internal battery backup

Physical Specifications

Dimensions (W x D x H)250mm x 156mm x 371mm
Weight7.8 kgs. (main unit with battery)

Environmental Requirements

Storage Temperature(-)10 - 45°C
Relative Humidity10 - 85%
Operation Temperature10 - 40°C

Options

Nebulizer
Humidifier
Stand
Carry Bag
Extended Battery
SpO2 Monitor
EtCO2 Monitor